EMIT II Plus 6-Acetylmorphine Assay

Accurately screen for heroin use

EMIT II Plus 6-Acetylmorphine Assay Performance

 Assay Principle  Homogeneous Enzyme Multiplied Immonoassay (EMIT)
 Cut-off  10 ng/mL
 Assay range  0-20 ng/mL
 Sample  Urine  
 Reagent Stability  Unopened until expiration date; Opened working reagent, 4 weeks
 Calibration Frequency  Based on control results

Qualitative Precision (mA/min)

LevelMeanRepeatabilityWithin-Lab
(ng/mL)SD%CVSD%CV
Control 17.5 ng/mL5581.50.371.3
Cut-off10.0 ng/mL5911.70.36.51.1
Control 212.5 ng/mL6221.70.361

Semi-quantitative Precision (ng/mL)

LevelMeanRepeatabilityWithin-Lab
(ng/mL)SD%CVSD%CV
Control 17.5 ng/mL8.70.121.40.435.1
Cut-off10.0 ng/mL11.30.141.50.484.2
Control 212.5 ng/mL14.20.161.10.53.5
lg

Assay Specificity

Concentrations of compounds that produce negative results vs. the 10 ng/mL cut-off

Compounds Concentration Tested(ng/mL)
Buprenorphine1,000,000
Codeine500,000
Dextromethorphan100,000
Dihydrocodeine500,000
Heroin80
Hydrocodone300,000
Hydromorphone100,000
Imipramine200,000
Levorphanol100,000
Meperidine800,000
Morphine100,000
Morphine-3-Glucuronide600,000
Morphine-6-Glucuronide600,000
Nalorphine100,000
Naloxone300,000
Naltrexone300,000
Norcodeine600,000
Normophine100,000
Oxycodoen400,000
Oxymorphone100,000

Ordering information

Bu web sayfasında gösterilen ürünler/özellikler her ülkede piyasada bulunmamaktadır. Mevzuata bağlı nedenlerle gelecekte bulunabileceği garanti edilemez. Lütfen daha fazla bilgi için yerel Siemens organizasyonuyla iletişime geçiniz.

Bu bilgi yardımcı oldu mu?
Yanıtınız için teşekkür ederiz
Yorumunuzda sayı ve/veya '@' işareti olduğunu fark ettik. Sizinle iletişime geçmemiz için telefon numarası ya da e-posta adresi girmeye çalışıyorsanız lütfen bizimle e-posta formumuz aracılığıyla iletişime geçiniz.

Not: Doğrudan kişisel bir yanıt almak için lütfen bizimle e-posta formumuz üzerinden iletişime geçiniz.

Teşekkür ederiz