Viva-Jr System

Gold-standard drug-testing performance with EMIT technology

 

Throughput
  • 66 EMIT® tests per hour
Reagent and sample system
  • One rotor combining both sample and reagent positions
  • Inner rotor ring:
    • 18 positions for reagent bottles
    • Cooled to approximately 8-12°C
  • Outer rotor ring:
    • 25 positions for samples, controls and calibrators
    • 3 positions for reagent bottles at ambient temperature
  • All reagent positions can be assigned as R1, R2 and R3
  • Reagent 1 volume 110-399 μL
  • Reagent 2 volume 0-298 μL
  • Reagent 3 volume 0-298 μL
  • Preheated reagent and sample needle with level detection and integrated mixer
  • Continuous sample loading
  • 13 mm OD primary tubes
  • All sample positions can contain 5 mL primary tubes or sample cups
  • Sample volume 2-30 μL per test, programmable in steps of 0.1 μL
Sample predilution Available in the 5 open channels, programmable ratios 1:5, 1:10, 1:20, 1:30, 1:40, 1:50 with 3 possible diluents
Pipetting system
  • KLOEHN syringes and valve block
  • Reagent syringe 1000 μL
  • Sample syringe 100 μL
Cuvette rotor
  • Semi-disposable rotor with 48 cuvettes, path length 7 mm
  • Minimum measuring volume 220 μL
  • Measuring temperature 37 °C controlled by Peltier elements
Washing units
  • Cuvette-washing with 4 x 500 μL of water:
    • The unit is equipped with liquid sensors
    • Waste can be collected in an external waste collection bottle, level float protected, or dispensed directly to a external drain
    • Cuvettes are dried before use
  • Distilled water: 1 L per hour while in use
Light source Quartz-iodine lamp 12 V - 20 W
Wavelength range Automatic wavelength selection by an 8-position filter wheel (340, 415, 505, 546, 570, 600, 660 and 700 nM). Half bandwidth 8 to 12 nM
Photometric range -0.1 to 3.0 Absorbance
Analytical modes
  • Kinetic measurement with linearity check
  • Bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
  • Two point measurement
  • Automatic rerun with sample reduction
Ambient temperature
  • 15 - 32°C
  • Maximum humidity 85%
Calculation modes
  • Logit Log
  • Non-linear calibration curves
  • Graphic plot of all measuring points
  • Prozone check for immunology tests
  • Cut-off declaration

 

 

Bu web sayfasında gösterilen ürünler/özellikler her ülkede piyasada bulunmamaktadır. Mevzuata bağlı nedenlerle gelecekte bulunabileceği garanti edilemez. Lütfen daha fazla bilgi için yerel Siemens organizasyonuyla iletişime geçiniz.

Bu bilgi yardımcı oldu mu?
Yanıtınız için teşekkür ederiz
Yorumunuzda sayı ve/veya '@' işareti olduğunu fark ettik. Sizinle iletişime geçmemiz için telefon numarası ya da e-posta adresi girmeye çalışıyorsanız lütfen bizimle e-posta formumuz aracılığıyla iletişime geçiniz.

Not: Doğrudan kişisel bir yanıt almak için lütfen bizimle e-posta formumuz üzerinden iletişime geçiniz.

Teşekkür ederiz