Dimensions
Main unit and cabinet
1858 x 1042 x 1415 mm (W x D x H) without LAS connection
2156 x 1042 x 1415 mm (W x D x H) with LAS connection
 

Measurement channels
25 channels total for clotting, chromogenic, and immunoassay: 4 with stirring for opto-mechanical and aggregation methods, 1 dedicated channel for LOCI assays
 

Sample capacity
150 samples, 30 racks with five samples per rack (maximum load)
 

Throughput (tests/hour)7
PT/APTT 210 (single-parameter analysis)
PT and APTT 350 (simultaneous analysis)
PT/APTT/AT 330 (simultaneous analysis)
PT/APTT/AT/DD 310 (simultaneous analysis)
 

Operation

  • Access mode: Continuous random access (samples, cuvettes, reagents, controls and calibrators, liquid waste, water)
  • STAT: STAT sample rack with five priority positions. Automatic prioritization
     

Reaction tubes

  • Type: Single reaction cuvettes
  • Loading: Automatic continuous access; 1500 cuvettes onboard
     

Data storage
Archive viewer: 5000 results maximum per archive set, number of archive sets limited only by disc space
Job list: number of samples and results limited only by disc space