Tools


Site Haritası

Site Haritası

IgG Subclass 1–4 Immunoassays

The Comprehensive Solution for IgG Determination


Assay characteristics

 N Antiserum IgG/Subclass 1N Antiserum IgG/Subclass 2N Latex IgG3N Latex IgG4
Assay principleLatex-enhanced immunonephelometry
Sample
types
Human serum, EDTA plasma, heparinized plasma
Reference ranges*4.05 – 10.111.69 – 7.860.11 – 0.850.03 – 2.01
Initial measuring range0.85 - 270.35 - 110.033 - 2.10.052 - 3.3
Measuring time6122+6
incl. pre-reaction
2+6
incl. pre-reaction
Once-opened reagent staibilty2 weeks on BN ProSpec® System, 3 days (6 days with evaporation stoppers) on BN™ II System2 weeks on BN ProSpec® System, 3 days (6 days with evaporation stoppers) on BN™ II System2 weeks on BN ProSpec® System, 5 days (10 days with evaporation stoppers) on BN™ II System2 weeks on BN ProSpec® System, 5 days (10 days with evaporation stoppers) on BN™ II System
Calibration frequency4 weeks4 weeks4 weeks4 weeks
Precision
(total CV)
≤ 3% @ 3.6 g/L< 3.0% @ 1.8 g/L≤ 6% @ 0.84 g/L< 3% @ 0.507 g/L

 

*The above reference ranges (2.5th to 97.5th percentile in g/L) were established by testing samples from 279 apparently healthy adults from Central Europe (data on file). Please refer to the download section for pediatric reference ranges.

Bu web sayfasında gösterilen ürünler/özellikler her ülkede piyasada bulunmamaktadır. Mevzuata bağlı nedenlerle gelecekte bulunabileceği garanti edilemez. Lütfen daha fazla bilgi için yerel Siemens organizasyonuyla iletişime geçiniz.

Bu bilgi yardımcı oldu mu?
Yanıtınız için teşekkür ederiz
Yorumunuzda sayı ve/veya '@' işareti olduğunu fark ettik. Sizinle iletişime geçmemiz için telefon numarası ya da e-posta adresi girmeye çalışıyorsanız lütfen bizimle e-posta formumuz aracılığıyla iletişime geçiniz.

Not: Doğrudan kişisel bir yanıt almak için lütfen bizimle e-posta formumuz üzerinden iletişime geçiniz.

Teşekkür ederiz