BN II System

Confidence in plasma protein results

Measuring principle Nephelometry; measurement of the scattered light intensity in a fixed angle of 13 - 24 degrees
Methods More than 60 programmed assay protocols
Sample throughput Effective: Approx. 130 tests/hour depending on the assay mix
Nominal: 225 tests/hour
Analysis method Fixed-time kinetics, end-point measurement, VLin Integral
Calibration Multi-point calibration
Rack transport unit Racks for up to 8 standard or control vials
Racks for up to 7 reagent vials
Racks for up to 10 sample tubes
Dilution unit 2 frames for max. 264 dilution cups
Size of sample tubes Diameter: 12 - 16 mm
Height: 55 - 100 mm
For pediatric samples:
conical microtubes with a maximum filling volume of 1.5 mL
Barcode types Automatic reading of different barcode types:
2/5 interleaved
Codabar
Code 39
Code 93
Code 128
UPC
EAN
Reagent volume 40 µL reagent consumption on average
Sample dilution 1 : 1 to 1 : 40,000
Level detection For samples, standards, controls, reagents, and system liquid containers
Reaction cuvettes 60 reusable cuvettes
Measuring temperature 37 ± 1.5°C
Light source Infrared high performance LED
Wavelength 840 ± 25 nm
Detector Photodiode with integrated pre-amplifier
Instrument weight and dimensions Analyzer dimension (width x height with tubings x depth)
124.5 cm x 91.4 cm x 63.5 cm (49 in. x 36 in. x 25 in.)
Computer dimensions (width x height x depth)
35.6 cm x 50.8 cm x 55.9 cm (14 in. x 20 in. x 22 in.)
Depth with keyboard is 73.7 cm (29 in.)
Printer dimensions (width x height x depth) 
40.6 cm x 266 cm x 36.8 cm (16 in. x 10.5 in. x 14.5 in.) 
Total bench length 
226.1 cm (89 in.) 
Weight 
Analyzer: 150 kg (330 lbs.) 
Terminal: 6.4 kg (14.08 lbs.)
Room environment Ambient temperature 
15 to 32 °C (59 to 89 °F) 
Relative humidity 
Maintain between 30% and 85% (noncondensing) 
Average thermal output 
340 BTU/hr (100 W) in standby mode 
975 BTU/hr (286 W) in operating mode 
Water requirements 
Deionized water, NCClS Type 2 
Microbial count must not exceed 100 CFU/ml 
Electrical installation requirements Voltage 
100 to 127 V/50 or 60 Hz 
187 to 240 V/50 or 60 H. 
Power consumption 
<140 VA (stand-by-mode) 
<400 VA (operating mode) 
Host interfacing 
Download and host-query mode 
Computer and Printer Terminal 
Macintosh PC 
Printer 
Lexmark E260 D or DN or equivalent

Bu web sayfasında gösterilen ürünler/özellikler her ülkede piyasada bulunmamaktadır. Mevzuata bağlı nedenlerle gelecekte bulunabileceği garanti edilemez. Lütfen daha fazla bilgi için yerel Siemens organizasyonuyla iletişime geçiniz.

Bu bilgi yardımcı oldu mu?
Yanıtınız için teşekkür ederiz
Yorumunuzda sayı ve/veya '@' işareti olduğunu fark ettik. Sizinle iletişime geçmemiz için telefon numarası ya da e-posta adresi girmeye çalışıyorsanız lütfen bizimle e-posta formumuz aracılığıyla iletişime geçiniz.

Not: Doğrudan kişisel bir yanıt almak için lütfen bizimle e-posta formumuz üzerinden iletişime geçiniz.

Teşekkür ederiz