Measuring principle Nephelometry; measurement of the scattered light intensity in a fixed angle of 13 - 24 degrees
Methods More than 60 programmed assay protocols
Sample throughput Effective: Approx. 65 tests/hour depending on the assay mix
Nominal: 100 tests/hour
Analysis method Fixed-time kinetics, end-point measurement, VLin Integral
Calibration Multi-point calibration
Rack transport unit Segments for 3 control serum vials
Segments for 2 reagent vials
Segments for 15 sample tubes
Dilution unit 1 frame for max. 132 dilution cups
Size of sample tubes Diameter: 11 - 16 mm
Height: 55 - 100 mm
For pediatric samples:
conical microtubes with a maximum filling volume of 1.5 mL
Barcode types Automatic reading of different barcode types:
2/5 interleaved
Codabar
Code 39
Code 93
Code 128
Reagent volume 40 µL reagent consumption on average
Sample dilution 1 : 1 to 1 : 32,000
Level detection For samples, standards, controls, and reagents
Reaction cuvettes 90 disposable cuvettes
Measuring temperature 37 ± 1.5°C
Light source Infrared high performance LED
Wavelength 840 ± 25 nm
Detector Photodiode with integrated pre-amplifier
Instrument weight and dimensions Analyzer dimensions (width x height x depth)
109.2 cm 81.3 cm x 63.5 cm (43 in. x 32 in. x 25 in.)
Computer and monitor dimensions (width x height x depth incl. keyboard)
36.2 cm x 45.7 cm x 54.6 cm (14.25 in. x 18 in. x 21.5 in.)
Printer dimensions (width x height x depth)
40.6 cm x 26.6 cm x 36.8 cm (16 in. x 10.5 in. x 14.5 in.)
Weight
Analyzer: 115,2 kg (254 lbs.)
Computer: 5.9 kg (13 lbs.)
Monitor: 12.6 kg (27.7 lbs.)
Printer: 7.3 kg (16.1 lbs.)
Room environment Ambient temperature
18 to 32 °C (64 to 89 °F)
Relative humidity
Maintain between 30% and 85% (noncondensing)
Average thermal output
340 BTU/hr (100 W) in standby mode
975 BTU/hr (286 W) in operating mode
Water requirements
Deionized water NCClS Type 2
Microbial count must not exceed 100 CFU/ml
Electrical installation requirements Voltage
117 to 127 V/50 or 60 Hz
207 to 253 V/50 or 60 Hz
Power consumption
<140 VA (standby mode)
<400 VA (operating mode)
Host interfacing Download mode, host-query mode, ASTM
Ports 1 Serial (DB-9) (male)
1 10-Base-T/Base TX Ethernet RJ45
Computer and Printer Terminal
Windows® XP
Printer
Lexmark E260 D or DN or equivalent

Bu web sayfasında gösterilen ürünler/özellikler her ülkede piyasada bulunmamaktadır. Mevzuata bağlı nedenlerle gelecekte bulunabileceği garanti edilemez. Lütfen daha fazla bilgi için yerel Siemens organizasyonuyla iletişime geçiniz.

Bu bilgi yardımcı oldu mu?
Yanıtınız için teşekkür ederiz
Yorumunuzda sayı ve/veya '@' işareti olduğunu fark ettik. Sizinle iletişime geçmemiz için telefon numarası ya da e-posta adresi girmeye çalışıyorsanız lütfen bizimle e-posta formumuz aracılığıyla iletişime geçiniz.

Not: Doğrudan kişisel bir yanıt almak için lütfen bizimle e-posta formumuz üzerinden iletişime geçiniz.

Teşekkür ederiz